| Use Restrictions | |
| EMC Description | EMC: Class A products, see the manufacturer's declaration in the download area |
| Product Type | temperature measurement transmitter |
| product series | MINI Analog |
| Channels | 1 |
| configuration | DIP switch |
| insulation characteristics | |
| Overvoltage Category | II |
| pollution level | 2 |
| electrical isolation | 3 End Isolation |
| rated conditions | 235.5 mW |
| protection circuit | transient protection |
| step response (0-99%) | < 160 ms |
| maximum temperature coefficient | < 0.02 %/K |
| set range | ((100 K/set measuring temperature range [K]) +0.1)% |
| Full-scale transmission fault | ≤ 0,2% |
| Electrical isolation input/output/power supply | |
| Rated insulation voltage | 50V AC/DC |
| Test Voltage | 1.5 kV AC (50 Hz, 60 s) |
| insulation | foundation isolation conforming to IEC/EN 61010 |
| Power | |
| Rated supply voltage | 24V DC |
| Supply Voltage Range | 19.2V DC ... 30V DC (DIN rail bus connector (ME 6,2 TBUS-2 1,5/5-ST-3,81 GN, order number 2869728) can be used to bridge the power supply. Can be stuck to a 35mm DIN rail conforming to the EN60715 standard)) |
| Maximum current consumption | < 21 mA (24V DC) |
| power consumption | < 500 mW |
| signal | |
| Input Number | 1 |
| measurement | |
| configurable/programmable | is |
| Available Sensor Types (RTDs) | Pt 100(IEC 60751/EN 60751) |
| Temperature measurement range | minimum 50 K |
| Sensor Type: | -150°C ... 850°C (configurable) |
| Sensor Input Current | 1 mA (constant) |
| Allowable total conductor resistance | 10Ω (per cable) |
| connection system | 2,3,4 lines |
| signal: current/voltage | |
| outputs | 1 |
| configurable/programmable | Yes, not configured |
| voltage output signal | 0V ... 5V |
| 1V ... 5V | |
| 0V ... 10V | |
| 10V ... 0V | |
| Maximum voltage output signal | 12.5V |
| No-load voltage | 12.5V |
| current output signal | 0 mA ... 20 mA |
| 4 mA ... 20 mA | |
| 20 mA ... 0 mA | |
| 20 mA ... 4 mA | |
| Maximum current output signal | 23 mA |
| short-circuit current | 10 mA |
| load/output load voltage output | ≥ 10 kΩ |
| load/output load current output | < 500 Ω (when 20 mA) |
| fluctuation | < 20 mVSS(at 10 kΩ) |
| < 20 mVSS(at 500 Ω) | |
| Connection Mode | screw connection |
| stripping length | 12mm |
| thread | M3 |
| Rigid conductor cross section | 0.2 mm² ... 2.5 mm² |
| Flexible Wire Cross Section | 0.2 mm² ... 2.5 mm² |
| Attachable Guide Cross Section AWG | 26 ... 12 |
Phoenix MINI MCR-SL-PT100-UI-NC-thermal resistance measuring transmitter 2864273